    User name Solvignon

    Log entry time 12:37:40 on January21,2003

    Entry number 92804

    keyword=EPR tests with gradient coils on

    The target is in empty target position for beam study.
    The gradient coils are ramped to 5 A each, 0.5 A per minute.

    In the following:
    Tgc= Transverse gradient coils; Lgc= Longitudinal gradient coils.

    1) Tgc= 5A, Lgc =5A (figure #1)

    dn: P=(36.38 +/- 0.29)%
    up: P=(35.02 +/- 0.28)%

    2) Tgc= 5A, Lgc =4A

    dn: P=(35.40 +/- 0.32)%
    up: P=(33.90 +/- 0.30)%

    holding field at 24.38 G

    3) Tgc= 5A, Lgc =3A

    holding field at 24.35 G

    dn: P=(34.63 +/- 0.30)%
    up: P=(33.51 +/- 0.30)%

    holding field at 24.34 G

    4) Tgc= 6A, Lgc =3A (figure #2)

    holding field at 24.33 G

    dn: P=(33.98 +/- 0.35)%
    up: P=(33.19 +/- 0.39)%

    holding field at 24.33 G

    5) Tgc= 6A, Lgc =2A

    holding field at 24.32 G

    dn: P=(34.12 +/- 0.28)%
    up: P=(32.03 +/- 0.48)%

    holding field at 24.31 G


    From this test, EPR is more stable for settings #4.
    But it's also the only measurement with the field stable.

    I stopped the tests for now because the polarization dropped
    at 32% but I will need to do more tests during the next long
    down time.


    FIGURE 1

    FIGURE 2