    Log entry time 06:24:04 on January22,2003

    Entry number 92962

    This entry is a followup to: 92952

    keyword=DAQ hangs again


    Another DAQ hang, same symptoms (ts1 hangs, event collection stops of L-HRS DAQ).

    Note that ts1 is the trouble-maker, not ts0 as reported earlier.

    L-HRS CODA killed and restarted.


    We noticed that rcServer was running on adaqs2 on the R-HRS console also, so
    there was no rcServer on adaql2. In concern that we had started the DAQ
    incorrectly, we restarted both L and R DAQs.

    Each time we restart the L-HRS DAQ, the download step hangs on the first try
    while talking to EB1. After the download fails (or is cancelled), and a reset is done,
    the download works. This happens consistently when restarting the L DAQ, even
    though we follow the Configure-Reset-Download voodoo ritual.