• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT

    User name R. Michaels

    Log entry time 22:32:26 on January23,2003

    Entry number 93144

    Followups:

    keyword=ped run L-arm, new pedestals

    pedrunL config doesn't work, they tell me. Hard to fix
    from home, so I'll look tomorrow (It was working at start
    of experiment...). But one can also get pedestals by hand.
    In either adaqs2:~adev or adaql2:~a-onl (for L-arm, R-arm
    resp.) do the following (stuff in quotes are commands) :

    "cd ~/ped"
    First you must load zero pedestals
    "loadzero.com"
    Now do an ordinary CODA run with ps8 = 2. And then:
    "ls /adaqs2/data1/e01012_20516.dat.0 > run.list"
    Run the analyzer (based on ped_auto.f)
    "./ped_auto_L > run.log" (yes, dot-slash-ped_auto_L)
    Look at the results in fort.19 and run.log
    A few failures and forced zeros are normal. If you like it, then
    "cp fort.21 ped3.dat"
    The file ped3.dat is loaded each run for ROC3 for L-arm.
    Similarly ped1.dat and ped2.dat on R-arm on adaql2:~/ped.
    I did it just now for L-arm. Looks ok.
    See also ~/ped/README.

    Unrelated issue: halog 93141 said there were event
    mismatches. According code in adaql2:~adaq/synch
    (see README as usual), there were indeed synch problems
    but this does not affect pedestal measurements. However
    it means you have another (different) disease. I guess
    I'll get paged again if it is found harmful. Gulp.