Main INDEX
Monthly INDEX
PREV
NEXT
User name R. Michaels
Log entry time 10:14:26 on February 4,2003
Entry number 94705
This entry is a followup to: 94625
keyword=re: HRS-L data 18:00-23:00
When we bypassed the RT module last night (from 18:00 to 23:00)
the triggers only had gates if scintillators had a hit. So T9 or T8 (pulsers)
had no gates. The fastbus would wait about 1 msec before giving up.
This is a small deadtime for the rates you ran, and not helicity correlated.
It explains why the zero helicity events were suppressed, since these come
in the vicinity of T9. The data are ok, I believe.