User name choi
Log entry time 18:01:10 on February10,2003
Entry number 95945
keyword=Helicity mode test
We have performed a series of helicity mode test. From online analysis, HRS DAQ agrees pretty well with HAPPEX DAQ. When large charge asymmetry has been generated, both of us are measuring 2.2% (HRS DAQ) and 2.4% (HAPPEX DAQ). The following table gives various conditions with corresponding run numbers.
Run Numbers
Conf. Pockel_Cell_V Delay Qrt/Pair Random/Toggle HRS HAPPEX MOLLER
===============================================================================
1 OFF 8 Qrt Random 916 1925 10579
2 OFF 0 Qrt Random 917 1926 10580
3 ON 0 Qrt Random 918 1927 10581
4 ON 0 Pair Random 919 1928 10582
5 OFF 0 Pair Random 920 1929 10583
6 OFF 0 Pair Toggle 921 1930 10584
7 ON 0 Pair Toggle 922 1931 10585
===============================================================================
Most of the settings are also recorded correctly in the end of run halog entry except Random/Toggle mode. In the end of run entry, it always says "OFF".