• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT

    User name Moffit

    Log entry time 17:24:51 on April22,2003

    Entry number 98084

    keyword=Parity: ADC Gain Study at 1 uA

    Note: Beam used during this time to check raster.
    RD = 0, IT= 13300, OS=1

    Run I (uA) What
    -----------------------------------------------
    2721 1 All CH ADCs in LoGain
    2722 1 All CH ADCs in HiGain


    Result: ADCs 23,13,12,14 saturate in HiGain (They should work fine at 1uA in LoGain).

    Will fix default so that the following ADCs will be in HiGain:
    2, 22, 15, 7
    While these ADCs will be in LoGain
    27, 23, 13, 12, 14
    (27 is used by Lumi)