    User name j. singh

    Log entry time 23:08:12 on April26,2003

    Entry number 98682

    Followups:

    keyword=trouble turning on septum

    22:40 attempted to turn on septum
    coil = 6.65K
    yoke = 4.80K
    He Level = 87%
    L1/L2 flow = 11.93
    all interlocks are green
    operating conditions look good

    turned on dump switch => light is green
    turned on power supply => light is red (stays this way for 10 min)

    called J LeRose
    recommends:
    1) flip power supply back to off
    2) flip dump switch back to off
    3) reset interlocks
    4) wait a while.
    5) go through procedure listed in manual & be patient