User name j. singh
Log entry time 23:08:12 on April26,2003
Entry number 98682
Followups:
keyword=trouble turning on septum
22:40 attempted to turn on septum
coil = 6.65K
yoke = 4.80K
He Level = 87%
L1/L2 flow = 11.93
all interlocks are green
operating conditions look good
turned on dump switch => light is green
turned on power supply => light is red (stays this way for 10 min)
called J LeRose
recommends:
1) flip power supply back to off
2) flip dump switch back to off
3) reset interlocks
4) wait a while.
5) go through procedure listed in manual & be patient