User name LeRose
Log entry time 00:19:54 on April28,2003
Entry number 99029
Followups:
keyword=Magnet Settings
Quads are back on and have been cycled. Septum set to Iset=152.8, Iout=168.75.
Septum settings determined from quad-off studies. SNAKE indicates that we expect to run out of acceptance (with quads off) at 25% above the correct septum setting. We observed the elastic peak essentially fading away at 75% above what we have heretofore called the "nominal" setting. Inspection of th_rot vs x_rot spectra for different septum settings suggests that Iset-152.8 is about right. Thin sieve has been installed and we shall see.
FIGURE 1
FIGURE 2