User name J. Singh
Log entry time 05:10:46 on May17,2003
Entry number 102301
keyword=Problem?:T1/T2 Rates for +3%
These were the T1/T2 rates for +3%:
run Target T1(Hz) T2(Hz)
2067 Empty 425 388
2068 Multi Foil 2423 1093
2069 Long Foil 1677 2071
2070 Long Foil 3293 517
2071 Multi Foil 3810 348
2072 Empty 1701 496
These rates look strange. The T2 is quite large and even bigger than T1 in some cases. After a discussion with Bogdan, the preliminary conclusion is that b/c +3% is at the edge of the acceptance, alot of the scattered electrons are missing the 2nd scintillator plane (S2). The T1/T2 rates for the 0% looked consistent with this idea. However, it's not obvious that we should see that high a rate in T2 for the +3%. See figs.1 and 2 for the focal plane plots.
FIGURE 1
FIGURE 2