    keyword=Polarization Loss Tests w/ Septum On

    During the Day shift, Tim Holstrom and I performed AFP-loss tests at 309 Amps on the Septum magnet. Below are the results.

    NMR-AFP Loss

    time up (mV) down (mV)
    11:12 17.81 17.76
    11:14 17.73 17.69
    11:16 17.62 17.58
    11:18 17.48 17.43
    11:20 17.35 17.30


    The average relative NMR AFP loss is 0.581%/measurement

    EPR-AFP Loss

    EPR Down (%) Up (%)
    1 38.1 37.4
    2 36.9 36.2
    3 35.8 35.4
    4 34.7 33.9
    5 33.8 33.3


    The average relative EPR AFP loss is 1.4%/sweep