User name Sulkosky
Log entry time 01:27:15 on July19,2003
Entry number 104639
keyword=Polarization Loss Tests w/ Septum On
During the Day shift, Tim Holstrom and I performed AFP-loss tests at 309 Amps on the Septum magnet. Below are the results.
NMR-AFP Loss
time up (mV) down (mV)
11:12 17.81 17.76
11:14 17.73 17.69
11:16 17.62 17.58
11:18 17.48 17.43
11:20 17.35 17.30
The average relative NMR AFP loss is 0.581%/measurement
EPR-AFP Loss
EPR Down (%) Up (%)
1 38.1 37.4
2 36.9 36.2
3 35.8 35.4
4 34.7 33.9
5 33.8 33.3
The average relative EPR AFP loss is 1.4%/sweep