Main INDEX
Monthly INDEX
PREV
NEXT
User name paschke
Log entry time 11:24:29 on August 1,2003
Entry number 108459
This entry is a followup to: 108433
keyword=New PC-HV (PITA) setpoints
For both IHWP IN and OUT:
PC (PITA) POS = 7.855
PC (PITA) NEG = 7.046
Details:
A second PITA scan was done with the IHWP inserted. The changes were
negligible:
DeltaV HV+ HV- Ev_low Ev_high
0 7.830 7.071 5200 9000
200 8.330 6.571 13000 16000
-200 7.330 7.571 17000 22000
This scan was done at 10 uA, so BCM2 was saturating and only BCM4 is
useful. Results are shown below. Final PC HV setpoints are only about
20 Volts from IHWP OUT values: PC-POS = 7.880, PC-NEG = 7.021.
I suggest only 1 set of Pockels cell HV setpoints, for simplicity.
For both IHWP IN and OUT:
PC (PITA) POS = 7.855
PC (PITA) NEG = 7.046
FIGURE 1