User name J. Singh
Log entry time 16:00:16 on August 2,2003
Entry number 108782
keyword=Septum LHe level dipped
1521 Noticed that the Septum liquid helium level dropped below the alarm level of 86%. Then it recovered about 6 mins later. Just to be on the safeside, we set the septum current to 0A and it ramped down slowly. We were worried about the septum quenching and warming up. Ramping back up on the septum should only take about 1/2 an hour. We were at a readback of 212.75A (wanted to be back at 212.66A)