keyword=curiosity in begin / end of run summary
it seems that some EPICS variables recorded in the run summaries differ between the beginning and the end. my best guess is that the following happens:
(1) the last run of a particular config ends
(2) the config is changed to the next setting
(3) the end of run summary script is still running during the config change
(4) the end of run summary, therefore, logs the EPICS variable value from during the config change and NOT from during the run
it is my belief that this was the situation that occured when the BHWP was changed 5 out of the last 10 times.
see for example the following runs:
begin end
3098 3098
3106 3106
3117 3117
3136 3136
3186 3186
it is not clear to me what can be done to fix this. luckily, there is a lot of redundancy when it comes to this stuff. i have rechecked the BHWP orientation for the elastic running @ 2.8 GeV/ 6 deg. i believe what's listed in the run summary is accurate.
(incidentally, Vince has noticed that sometimes the end of run summaries also indicate that the lasers are off when they are on.)