    User name Sulkosky

    Log entry time 12:23:10 on August 9,2003

    Entry number 109677

    This entry is a followup to: 109580

    Followups:

    keyword=Pick-up Coils Position

    The difference between the pick-up coil and Pol. 3He position is only 0.1 revolutions, which is only 0.04 mm. Since the two positions are so close the effect on the NMR measurement is neligible. So there is no need to move the target to the pick-up coil position. Since the two positions are so close the lifting motor controller does not distinguish between the two. So even if you move to the pick-up coils, the controller still associates the position with the Pol. 3He position. I will change the pick-up coil encoder position to be the same as the Pol. 3He position.