User name Eugene
Log entry time 15:42:18 on August14,2003
Entry number 110671
keyword=Moller results
Beam conditions: 1.148 (accel) , Wien angles 30.2o.
1/2 wavelength plate was OUT.
The beam polarization:
Pz= 75.8+/-0.2(stat)+/-3(syst)%. - no leak through correction.
Pz= 76.2+/-0.2(stat)+/-3(syst)+/-0.1%(leak)%. - with the leak through correction.
The leakage from Hall C, working at 50µA was about 0.4%,
at Hall A slit position of 14.22, att=600.
The septum heated up by 0.5K with a 0.2uA beam on the Moller target.