User name R. Michaels
Log entry time 03:24:58 on August25,2003
Entry number 113391
Followups:
keyword=He levels, called cryo., Q1 won't reset
Re: L-arm cryo. Q3 levels still read zero.
Q1 Helium went to zero then quickly recovered (minutes).
We tried but failed to reset & restore Q1.
We called Hall A tech (Gary), he advised to call cryo on-call.
Gary said the probes were in some cases near the top of a vessel,
so a reading of zero could mean 80% when the level drops.
MCC also suggested calling cryo on-call. We reached the cryotech
and he said Q1 glitch is probably instrumental (such a fast loss
and recovery is impossible). Meanwhile the cryo tech also
does not think Q3 level is really low -- flakey gauge probably.
So, the theory is 2 independently flakey gauges and no real
problem with helium (?????).