• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT

    User name A. Tobias

    Log entry time 07:36:53 on August27,2003

    Entry number 113628

    keyword=NMR: 44.9 % and EPR: ?


    ** This is the first NMR/EPR measurement since 13:00 yesterday **

    06:27 Start moving from Empty Target to Polarized 3He Target
    06:32 Ladder at Polarized 3He Target. Will wait for target chamber RTDs to stablize.
    06:40 Start NMR measurement.
    06:44 Complete NMR: 44.9%
    06:55 As per "Instructions for EPR when Septum Current is > 300A" we need to set the C and D correction coils to the following current:

    dBz/dy = I_read*0.1193-16.48

    I_coil(c) = (dBz/dy - 0.6667)/13.33

    For I_read=409.00A, I_coil(c)=+2.37412A

    06:59 Set I_coil(c)=+2.37412A
    07:00 Set I_coil(d)=-2.37412A
    07:01 Jaideep finds EPR resonance at about 12MHz and does EPR: anywhere between 35 to 45% (signal is so choppy that it seems essentially useless).
    07:04 Zero current to coil C
    07:05 Zero current to coil D
    07:08 Move back to Empty Target.



    FIGURE 1