    Entry number 115129

    keyword=A1/A2 spectra

    the adjustment of HV was done by using the scripts
    onlana.C, a1_calib.C, and a2_calib.C in the directory e94107/aero.
    The updated db are in the same directory.
    The single electron peak position adjusted to 100 (A1) and 50 (A2).
    The sum of all PMTs(asum_c) has a peak at 600 for A1 and 1000 for A2.
    These lead to 7.5 photo-electrons in A1 and ~ 20 in A2.



    FIGURE 1

    FIGURE 2