    User name Reitz

    Log entry time 11:52:58 on January27,2004

    Entry number 116878

    keyword=RICH DAQ tests

    runs 2179 ff are junk runs to test the RICH DAQ
    2179 was using the previous setup, and getting high rates with
    a very loose trigger
    2180 is the same, but the veto from the sequencers was remove
    2181 - 2187 were used to trace signals with the scope

    at the end one thing was changed:
    - the L1A/Busy was fed into a gate with 750ns width before going
    into the coinc-unit to form the rich trigger, therefore T3s coming
    later are ignored. (they should have been before, but the
    veto was removed a little too early, which may have caused some
    inefficiency at high rates)