User name Breuer
Log entry time 00:15:44 on January28,2004
Entry number 116942
keyword=Running shift summary 16:00-24:00
16:00
17:10 called MCC for status: doing Harp scans right now to
diagnose ongoing beam trip problem
18:03 beam ready, move in waterfall target
ask for rastered 60 uA beam
18:06 get 61 uA beam, start run 2193
after run modify prescalers for next run
18:18 start run 2194; notice that OTR "data valid: no"
informed operator who will work on it.
18:40 OTR gives valid results; but width in y~0.34mm and in
x ~ 0.2
decide to tolerate this since the cooldown period is soon
18:58 Rich HV tripped, stop run, reset HV, start new run
Harp Scan done at 17:16 (Elog entry 1192665
IHAIC12: sigma-x = 0.141, sigma-y = 0.099
We will not do a further Harp scan before recooling entry at 20:00.
BPM-B at -0.01 /-0.08 (19:30)
NOTE: left septum was set
183.50 instead of 183.31 A
19:59 start ramping down left septum; ARM and Ed Faults go
into Hall to recool septum
~21:45 beam permit, starting to ramp up magnet
~22:50 MCC askes for C target for harp scan, go to Carbon target
22:20: MCC is finished with Harp scans, moving waterfall
target back in
Harp scan results:
1C12
7.4699998330325 uA CW MODE (DC) (IBCxxxxCRCUR2)
Raw Data File: IHA1C12.01272004_22:08
x sig: 0.1445 +/- 0.0085
u sig: 0.0978 +/- 0.0063
y sig: 0.0292 +/- 0.0013
22:23: left septum ready, target ready, ask for 60 uA, rastered
22:27, after starting run 2198, asked MCC if they could reduce
width seen on OTR; values are same as before cooling shutdown
22:33: OTR: sigma-x= 0.125, sigma-y=0.220 (two days ago 0.07/0.16)
Operator volonteered to call optics expert for advice.
NOTE: OTR is right after HARP, but does not necessarily give
same reading, needs to be cross calibrated; Operator changed
a filter, which changed the readings from 0.21/0.33 to 0.13/0.22
without changes to the beam.
With the current settings, used for 60 uA beam on Waterfall target,
we get the following corresponding values:
OTR sigma x/y Harp sigma x/y
0.125 / 0.220 0.145 / 0.030
(yes, y is a factor of 7 different)
~22:48 Noticed that left septum magnet was not at full current
as requested (readback was ~178 A) ; stopped run;
reentered curent value; now PS current is 182.5,
Magnet current is 183.5 A.
23:54 beam unstable, MCC wants retuning, putting in Carbon target
~0005 move waterfall target back in; get beam
looks like septa IOC don't communicate "Keep alive status" red;
ask for beam off since we don't see temperature
~00:00 reboot coda; ROCS 12 and 16 needed rebooting;
runs:
2193 Waterfall,
2194 Waterfall, prescalers have been changed
2195 Waterfall, Rich tripped shortly before end of run
2196 Waterfall
2197 Waterfall start at 19:27
2198 Waterfall start at 22:26, 60 uA; left Septum current ~ 5A too low
2199 Waterfall start at 22:52
2200 Waterfall start at 23:21 run stopped acquiring ~23:40 without
visible error message; running at a couple