• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT

    User name R. Michaels

    Log entry time 10:00:37 on April14,2004

    Entry number 117898

    keyword=A1,A2 threshold (gain)

    Bob & Bodo: We adjusted the gain on A1-or signal used for trig T6.
    Bogdan had increased the HV, so we observed that 1 p.e. is 50 mV
    now (for gain=1). Adjust the gain to 0.2 (it had been 0.3) so that
    the 30 mV cut is 3 p.e. Meanwhile, A2-or is 30 mV (gain=1) so we
    left it at gain=1, as it was in previous running.