• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT

    User name markowitz

    Log entry time 11:49:58 on April17,2004

    Entry number 118310

    keyword=A2 calibration

    We have been trying to understand the number of photoelectrons seen by A1 and A2. Here, for all trigger types, we see that the 1 p.e. peak is seen at channel 50 (after subtracting pedestal). Using a calibration of 50 channels per photoelectron.

    The second plot indicates that for mixed particle types, the peak is at 14 photoelectrons (when incorrection fitting with a gaussian).


    FIGURE 1

    FIGURE 2