• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT

    User name jpchen

    Log entry time 14:10:08 on April22,2004

    Entry number 118873

    This entry is a followup to: 118850

    Followups:

    keyword=target GUI and alarm handler

    Cryotarget group has made some changes yesterday:
    1) There is only one version of GUI will be updated, which sets in the
    cryotarget group's disk. So the new procedure to start the GUI will
    be after login in as atarg, do 'cd $GUI', which gets you to the right GUI directory, then 'tgtgui' to start the control system.
    In the local PC (cryotarg in the counting house), a old version of GUI
    will be kept in GUI_LOCAL. In case a network (or other) problem
    provents one to get to the cryotarget group's GUI disk, one can still
    use the old GUI in the local disk, but it will not be the updated version.
    2) The alarm handler system has been changed to be MEDM version.

    So appearantly, the old GUI was run yesterday. It should be changed
    back to the new one (I have talked Bodo and he will do that). I have set all the alarm limits and they should be good for the duration of the hypernuclear running. Let me know if there is any new problem.