User name E.Chudakov
Log entry time 19:54:09 on May24,2004
Entry number 122766
keyword=Shower detectors calibration
I made the shower/preshower (R-arm) calibration. The energy/p peak
(sh+ps) is attached (p is measured in MeV/c). The RMS of the peak
is 5.3% at 1.57GeV average arm momentum.
Also, a crude calibration of Pion rejector (L-arm) is done using
pions. The common absolute scale is arbitrary at the moment. It
can be corrected after we have some electrons in the L-arm,
I updated the database
e04012/DB/20040521/
db_R.ps.dat
db_R.sh.dat
db_L.prl1.dat
db_L.prl2.dat