    Yi found that the broad missing mass resolution observed unitl now was due to a right-arm theta offset that was not in the optics database.

    The KIN1 (15cm) and KIN2 (4cm) data have therefore been re-analysed with this new offset in the database file; the plots are included below.

    The resolution (sigma) for the 4cm target is 1.25 MeV, and 2.30 MeV for the 15cm.