Main INDEX
Monthly INDEX
PREV
NEXT
User name J. Singh
Log entry time 01:58:24 on May26,2004
Entry number 123074
Followups:
keyword=Hall A OTR @ 1C12
the otr had a MaX PIX error. usually this is a gain problem, so i decreased the gain from 0.7 to 0.50. now it is saying "Mask Adjustment." i varied them all over the place, but the error never went away. the x width is 0.260 mm and the spec according to the run plan is < 0.240 mm. will investigate further...
FIGURE 1
FIGURE 2