User name LeRose
Log entry time 21:32:00 on June 2,2004
Entry number 124459
This entry is a followup to: 124447
Followups:
keyword=Left controls
Let me try again.
After the reboot do the following to get the left arm quickly back to 2.06 GeV/c settings:
select 6 degrees (at the top of the tools screen)
run Q2/3 UP TO 1600 (THE REBOOT SHUTS THEM OFF)
open the dipole power supply control screen
type 2.06 in the momentum set field on the tools screen and then quickly type 743 in the Iset field for the dipole (this will turn the "regulator" off.)
wait a few minutes, make sure the nmr is locked
turn the regulator back ON (from the nmr screen)
wait for things to settle in.