User name Seonho Choi
Log entry time 00:01:03 on June 8,2004
Entry number 124919
This entry is a followup to: 124913
keyword=Dipole reset procedure
1. Let the current go to zero.
2. Push "0 Amps" button under Hall A controls (dipole)
3. Wait for He level (greater than 60%, green indication)
4. Push "Reset" button
5. Check that all faults are cleared. (May need to push "reset" button several times, ~10 minutes waiting period)
6. Push "On" button.
7. Set the current to 100 A and verify that it is ramping properly.
8. Set the desired momentum+epsilon, wait a minute.
9. Set it at the desired momentum+zero.
10. wait
11. wait
12. and wait