keyword=Tests of profile scanners motion
We tested DAQ readings of scanner positions. Rough calibration
gives change of about 90 channels per 1mm displacement of scanner.
Conclusions:
1. rprofx and rprofy variables corresponding to X and Y positions
of scanner in right arm should be swapped.
2. Short axis potentiometer (X direction) in left arm works
erratically, with fluctuations corresponding to +-2mm in position. This potentiometer should be exchanged.
3. DC voltage on potentiometers should be lowered, now
it saturates ADCs (go from ~10V to 9V ?)
4. DV voltage shows some drift, <20 channels with a period
of about few hundred second. It should not be an issue.
5. Readings of Y position in left arm for a short time were
increasing with slower rate than they should. After some time
they came back to normal rate. We should watch behavior of
this potentiometer!
6. When left arm scanner moves in Y direction, its X readings
flactuate. Is this coupling related to contact problems in
damaged X potentiometer when scanner moves?