    User name R. Michaels

    Log entry time 10:56:06 on June10,2004

    Entry number 125236

    keyword=Profile Scanner in Spectr. DAQ

    I put the signals from the UMass/Smith Profile Scanner into the
    spectrometer DAQ. The amplitude goes into the 5th channel of
    what used to be the S1 discrim. This makes triggers T5, T6
    for R,L arm resp. It shows up in an ADC and TDC, as usual.
    The X, Y positions show up in scalers. They are in the 31st &
    32nd (last 2) channels of the normalization scaler and are ~1 MHz.
    With student's help (Kirsten & Darcy) I verified X,Y respond
    with motion. At first the X,Y was swapped on R-arm, so I swapped
    the cables at the scaler (only at the scaler). This probably means
    the cables are swapped an even number of times between the detector
    and the scaler.

    The libParity.so (analysis package) was updated to see this detector.

    Bottom line is: We are ready to align the profile scanner.