    User name Z. Meziani

    Log entry time 20:04:22 on June12,2004

    Entry number 125469

    keyword=Septa heating study

    Septa heating studies.

    There doesn't seem to be a correlation between the position of
    the beam on target and the heating of the septa magnets.

    With the beam at 30 uA we steered the beam
    at BPMB at 3 different positions in (X,Y)
    (0,0), (-4,0), (+4,0) and (0,+4) back to (0,0)
    with a rater at 1x1 mm^2. The results of the Septa
    coil temperatures for all steering configurations are consistent
    (see table).
    When there was no beam the temperature are consistently lower.
    We are still puzzled about the Septa heating source origin.





    FIGURE 1