    User name aniol

    Log entry time 23:20:04 on June15,2004

    Entry number 125902

    keyword=helium level in left septum is low

    The left septum helium level hovers around the trip off point. When it falls below 80% the interlock light goes on. This causes the current in the left septum to fall. During prod run 2173 we noticed the left hrs detector signal started to fall at about trig no. 72500. Once the red interlock light goes off on the helium level, go into the light blue set current field and hit return. This should start the magnet ramping up again. The current fell from 278.25A to 245A.