    User name Peter L

    Log entry time 13:30:50 on June26,2004

    Entry number 126967

    keyword=Septa Current Set Pints

    Here is a screen shot of the new set points.  The problem is that the
     Magnet Current readout is only good to .5 (aka it jumps fromm 174.75
     to 175.25 with no readouts in between.)  Therefore Here are the final setpoints:
    Right
    SetPoint:275.35
    Readout: 274.25
    
    Left
    SetPoint:275.50
    Readout: 275.25
    Note:  I adjusted the setpoints such that the readouts matched those
     calculated (right=274.04  Left=275.05) and watched as the readout
     fluctuated back and forth from say for left 274.75 and 275.25 back
     and forth meaning that the actual reading should be about 275.  Same
     for the right, the limits on the readout limit our accuracy.
    
    


    FIGURE 1