User name J. Cahoon
Log entry time 18:44:17 on June28,2004
Entry number 127270
keyword=Profile Scan Runs 2509 & 2510
High Voltages for the profile scanner detectors were both set to -1400V. This seems to be a fair value to keep them at, since they were both saturating when I tested them at -2800 and -2000.
Left side's potentiometers seem to be working pretty good now.
Run 2509 -
Took a screenshot of all the settings, it can be found with the other screenshots on the motion controller computer. This run will have data for the right arm only since I miss-positioned the scan in the left arm by assuming it to be in the same place as the right arm.
Run 2510 -
Took a scan of the full possible area so that I can find out where the left stripe is located. Seems to be a little bit behind the right side. This means we're going to be able to see the whole thing!