    User name R. Michaels

    Log entry time 09:49:14 on June29,2004

    Entry number 127360

    Followups:

    keyword=Q2 measurement

    Q2 measurement.

    Run 1199 was a spot++ check. Looked ok.

    VDCs still off.
    1 uA on target.
    Rates:
    L-arm HAPPEX detector PMT 1 = 228 kHz
    L-arm HAPPEX detector PMT 2 = 358 kHz
    R-arm HAPPEX detector PMT 1 = 500 kHz
    R-arm HAPPEX detector PMT 2 = 160 kHz
    T1 = 730 kHz
    T2 = 620 kHz
    T3 = 670 kHz
    T4 = 520 kHz
    T5 = T6 = 0
    Run 1200 to log these rates.

    Next, we turn down the beam current to 0.2 uA and turn on VDCs

    Run 1201 for Q2 measurement
    T1 = 175 kHz
    T2 = 140 kHz
    T3 = 155 kHz
    T4 = 120 kHz

    Run 1202 is a repeat Q2.

    The data looked good online (detectors, tracks, etc).

    Realized dithering was still on. We turn it off and do run 1203.
    To some extent this "accident" (of running with dithering on)
    was not bad, since we run the experiment with dithering on,
    and we can verify it makes little difference (not expected to).

    Run 1203 -- without dithering.

    Fig 1 -- Run 1202   Q2 and missing mass.
    Required a hit in HAPPEX detector (either segment) taking
    into account the different gains.
    The most important thing for now is the SHAPE, and NOT THE VALUE.
    (Database is slightly wrong, e.g. I don't have surveyed angles).

    I checked the same plots for run 1203 (dithering off).
    Q2 looked the same but shifted lower
    by 0.08% and 0.12% on Left, Right HRS resp. (in the noise).



    FIGURE 1