User name fieschi
Log entry time 08:09:21 on October 3,2004
Entry number 132777
keyword=shift summary
00:00 start run 3031
00:35 end run 3031
00:36 start run 3032
01:11 end run 3032
01:12 bleedthrough measurement : 2.14%
01:15 start run 3033
01:50 end run 3033
01:53 start run 3034
02:26 end run 3034
02:33 start run 3035
02:40 end run 2035
02:49 start run 3036
daq problems :
- 7 bloc of the proton array have no signal :
channel 38 39 60 61 62 63 64
- channel 11 and 26 is read but the ADC values look like pedestal
call Alex to fix it
this problem appeared from run 3024 to run 3035
Alex fixed the first problem :
for the run 3036 blocs 38 39 60 61 62 63 64 are read
for bloc 11 an ARS problem is suspected : access needed to test
for bloc 26 need to check the gain with a higher HV value
03:40 end run 3036
change HV of the bloc 26 of the proton array
old HV value : -582V
new HV value : -750V
03:40 start run 3037
03:46 end run 3037
problem on bloc 26 is still here
04:11 call MCC to have an access to check the proton array detector
the HV channel on the crate was wrong : channel 26 is now plugged on a spare HV channel (cf dvcs ELOG entry )
the ARS slot for the channel 11 seems to be wrong. This card need to be changed.
channel 11 of the proton array is plugged on a spare slot of the ARS (cf halog entry)
06:30 back to beam permit :2.5uA
start run 3046
channel 26 works properly (cf dvcs elog entry)
we checked that this channel worked during older runs run 2881
03:40 end run 3036
06:32 start run 3046
07:31 end run 3046
problem of daq for runs 3046 and 3047
07:33 start run 3048
07:39 bleedthrough measurement :1.4%
run list
------------------
DVCS
-run 3031
-run 3032
-run 3033
-run 3034
-run 3035
-run 3036
-run 3037
-run 3046
-run 3047