    User name A. Camsonne

    Log entry time 17:15:40 on October 5,2004

    Entry number 133068

    keyword=DAQ stuffs during acess

    During the access I mostly took care of the LED and cosmics proton array trigger. I had to move the stop from the logic module to the calorimeter stop output in order to have these triggers work properly.
    So timing of the proton array might shift a bit ( transit time in trigger + cable ) but no more than 30 ns so I will not touch the ND parameter for the ARS for now. Cosmics trigger seems to work, this trigger goes to the input 7 of the trigger superviser. So cosmics or LED can be taken using DVCSCosmics or DVCSLed and putting ps7 to 1
    I also added a scaler with the 105 Khz clock in the proton array crate.