Main INDEX
Monthly INDEX
PREV
NEXT
User name Lindgren
Log entry time 12:04:21 on January 9,2005
Entry number 138692
keyword=Mid Shift Summary
8:00 AM - 9:45 AM Left Spectrometer stuck when trying to move it to 14.1 degrees. After restarting several times, it finally moved (When Doug arrived).
8:45 - 9:00 Run 1585 250K
9:01 - 9:13 Run 1586 250K
9:17 - 9:28 Run 1587 250K
9:32 - 9:44 Run 1588 250K
9:45 Change to tilted C target
recalculated prescale factors
ps3 = 100i ps6 = 2000
Noticed large 56% deadtime Run 1589.
Killed run. Started again.
Increased prescale factors to 200 and 4000. Deadtime reduced to 1%
Put ps3 back to 100, ps6 =4000 Run 1590. Low deadtime
Put prescale factors back to 100 and 2000. Still Low deadtime
Deadtime is okay now. We don't know what was wrong,
Start again.
10:18 - 10:31 Run 1592 251K
10:32 - 10:47 Run 1593 250K
Moved spectrometer to 15.5 degrees and changed field to 1.14720
11:12 - 11:26 Run 1595 251K
11:27 MCC requested beam for 20 minutes for injector studies.
11:58 Beam is Back
11:59 - Run 1596