User name Lindgren
Log entry time 16:21:27 on January 9,2005
Entry number 138716
keyword=Shift Summary
8:00 AM - 9:45 AM Left Spectrometer stuck when trying to move it to 14.1 degrees. After restarting several times, it finally moved (When Doug arrived).
8:45 - 9:00 Run 1585 250K
9:01 - 9:13 Run 1586 250K
9:17 - 9:28 Run 1587 250K
9:32 - 9:44 Run 1588 250K
9:45 Change to tilted C target
recalculated prescale factors
ps3 = 100i ps6 = 2000
Noticed large 56% deadtime Run 1589.
Killed run. Started again.
Increased prescale factors to 200 and 4000. Deadtime reduced to 1%
Put ps3 back to 100, ps6 =4000 Run 1590. Low deadtime
Put prescale factors back to 100 and 2000. Still Low deadtime
Deadtime is okay now. We don't know what was wrong,
Start again.
10:18 - 10:31 Run 1592 251K
10:32 - 10:47 Run 1593 250K
Moved spectrometer to 15.5 degrees and changed field to 1.14720
11:12 - 11:26 Run 1595 251K
11:27 MCC requested beam for 20 minutes for injector studies.
11:58 Beam is Back
11:59 - 12:10 Run 1596 251K
Changed target to 4 cm liquid hydrogen
12:28 - 12;40 Run 1597 251K
12:41 - 12:55 Run 1298 251K
Move Spectrometer to 17.0 degrees
Changed field to 1.13740
Set Prescale PS3. PS6 is the same.
13:16 - 13;29 Run 1599 251K
13:30 - 13:43 Run 1600 251K
Changed to tlted carbon target
13:56 - 14:09 Run 1601 251K
14:10 - 14:26 Run 1602 251K
14:41 - 14:52 Run 1603 251K
15:35 No beam for 10 minutes
Beam Back
Turn on Neutron Veto counters high voltage.
Set ps7 =10000 to chack ADCS
Left Vetos ADCs 0-7 not working
Right Vetos ADCs 0-7 not working
Turn High Voltage for Vetoes off
Start taking data again
Changed target to 4 cm LD2
Cycle Quads Q2 and Q3
Set Dipole to 1.152 GeV
Move LHRS to 21 degrees
Started Run 1604 250K
16:09 Start Run 1605