User name P. Monaghan
Log entry time 00:03:29 on January 18, 2005
Entry number 139542
This entry is a followup to: 139450
keyword=Neutron detector channel N1-27L
During the access this evening, I looked at the signal from the neutron
detector channel N1-27L which had a low scaler value and a small adc
spectrum. On an oscilloscope, I could see the signal coming out of the
amplifier and out of the discriminator, although the trigger rate on the
oscilloscope was slow. I compared the raw signal from this PMT to another
channel (N1-16L) and found that the signal from N1-27L was weaker. It
appears to be a problem with the pmt on this channel. Simply the high
voltage may need increasing a little bit, although at the moment it is
set to -1350 V and is comparable to other channels. Bodo believes at the
moment, we might be better to wait and go through the gain setting
process again for all channels, since there are a number that appear to
have a low scaler value.