    User name Saha

    Log entry time 09:45:57 on January27,2005

    Entry number 140162

    Followups:

    keyword=Raster Size - Beam Studies

    Raster Size - Beam Studies

    All widths are FWHM in mm.
    All quads between raster and Target position zeroed.
    Beam Current set to 5 uA.
    No Target.
    Harp Scans performed by MCC.

    CODA#, Rast. cur., HarpScan#, Spot++: Ax, Ay, Bx, By, Tarx, Tary

    2118, 0 amps, 0.08, 0.10, 0.08, 0.09, 0.11, 0.12
    2120, 9.9 amps, 1.1, 1.1, 1.5, 1.45, 1.65, 1.6
    2121, 20.3 amps, 2.7, 2.3, 3.6, 2.8, 4.0, 3.2
    2122, 30.8 amps, 3.3, 3.3, 4.3, 4.2, 4.9, 4.6
    2123, 40.6 amps, 4.1, 4.2, 5.7, 5.4, 6.4, 6.2