    User name Markowitz

    Log entry time 12:06:50 on June 06, 2005

    Entry number 145306

    keyword=changed software A1, A2 gains in online database directory

    I set the software gains all to 1.000 in the on-line database directory files (db_L.a1.dat and db_L.a2.dat). Armando, Bogdan and I have tried to set the HV such that there are 100 channels per photoelectron for all phototubes (for trigger purposes).

    The gains were about .5 but varied, meaning they have not been calibrated in software (yet) and give a mis-leading (and I hope worrisome) impression to a casual glance.

    Software gains do need to be tuned but should be accurate to 5% now. Armando and I can work on this further today.