User name E. Cisbani
Log entry time 15:56:14 on June19,2005
Entry number 146620
keyword=Shift Summary (Day 06/19)
Shift Summary
08:00-08:50 Trouble rebooting adaql2 (see Pete e-log 146570)
08:55-09:30 Run 4701 (H2O flux=749.8)
09:31-10:07 Run 4702 (H2O flux=750.1) no beam at the beginning for a couple of minutes
10:08-10:36 Run 4703 (H2O flux=743.2)
10:33-11:17 Run 4704 (H2O flux=749.0)
11:17-11:50 Run 4705 (H2O flux=748.3)
11:50-12:30 Run 4706 (H2O flux=746.5) Right-Septum tripped (around 12:00!), run stopped at 130k events
12:31 Right Septum reset, when ramped-up tripped again (Upper coil over temperature alarm)
12:50 PC called, he notice that the upper coil temperature is above 7.9 (upper limit). Suggest to switch off the beam in Hall A in order to speed up the cool down of the upper coil.
13:10-13:20 Run 4707 RICH pedestal run, beam off
13:22 Beam back at 55 uA
13:27-13:30 Run 4708 (H2O flux=744.2); yunk run (forgot to re-configure RICHNF1)
13:30-14:03 Run 4709 (H2O flux=751.3)
14:03-14:25 Run 4710 (H2O flux=750.6)
14:25 Problem with vacuum in the beam line. MCC turned off the beam and made a controlled access to restart the pump and check the gauge
They will call us when the beam is available.
14:30 Noticed that Right Septum quenched about 10 minutes ago.
Alarm reset, magnet ramped up
14:45 Pete suggest to reduce the pump speed at 600 (tacheometer)
14:55 Beam back at 55 uA
15:05-15:46 Run 4711 (H2O flux=600.2)
15:46-15:50 Run 4712 (H2O flux=607.6)
15:50 MCC informed about a new short controlled access to reset the same pump of the previous vacuum problem