    keyword=cavity commissioning

    J. Musson, A. Freyberger, R. Michaels, L. Kaufman
    
    Commissioning of cavity monitors near target (4A, 4B).
    14:42   Take out target.  Turn off raster.
    
    Time .... Spectr.Run ..... HAPPEX Run .... comments
    
    14:52 .... 2327 ..... raster on at start.
    14:55 .... 2328 .......... 3946 ..... Y scan
    Oops !! dithering was left on.  We will turn it off now.
    Note, the X,Y phases have not been adjusted yet.
    That's what Musson is trying to do now.
    
    14:57 .... 2329 .......... 3947 ..... Y scan
    Lot's of stuff in electronics was changing ...
    
    15:33 .... 2330 .......... 3948 .... more scanning.
    Musson says downstream cavity is anti-correlated to striplines in Y.
    We can fix it with access (put a 15 cm cable on one cavity output),
    but we'll wait on that; meanwhile simply multiply by "-1".
    
    15:41 .... 2331 .......... 3949 ..... X scan (mostly)
    Lots of stuff changing - gains, phase, ...
    
    It's realized that the raster is on, about 2A in X & Y,
    and we cannot turn it off.  The manual control unit in Hall A
    counting room is at zero but there's a nonzero field in the 
    readback (box on scope, not collapsed to a point), and spot++ 
    says the RMS is ~170 um in X & Y. 
    
    More scans to avoid the middle where signal "chatters"
    due to motion.
    
    16:38 .... 2232 ......... 3950 ..... more scans.
    
    Phase and gains are set now.  Commissioning test plan finished.
    Next, we do an X,Y scan to check things out.
    
    16:50 .... 2333 ......... 3951 ...... Scan X & Y for tests.
    (Spectr. DAQ: Horiz. up to event 26500, then vertical).
    
    We are done !  
    We may want to do a "bull's eye" scan at some point ... next week, 
    or someday.