User name lkaufman
Log entry time 00:02:47 on September 24, 2005
Entry number 155747
keyword=Large Aq scans for detector linearity (Runs 5315 and 5316)
Took two scans designed to generate large charge asymmetry, in order
to better test detector non-linearity. Most methods for generating
charge asymmetry also generates position differences, but we'll have
to live with that.
Each scan covered approximately 1/2 oscillation (peak-to-trough) of
the 4-theta term in the halfwave scan. A PITA offset delta = -180 V
was used to turn this into a significant charge asymmetry change.
This was done for the OUT halfwave plate state for 2 different beam
currents. The IA cell was set to 5.03 V. Feedback and dithering were
turned off.Hall C was off during these scans.
The auto_rhwp macro was used, with steps of 5 degrees and 180 sec per
point. Scan clean will point to good data at stable set points.
Run 5315: IHWP OUT, RHWP scan [10,60], PC Pos = 4.75, Neg = 6.81, beam
current = 37 uA
Run 5316: IHWP OUT, RHWP scan [10,60], PC Pos = 4.75, Neg = 6.81, beam
current = 25 uA