    User name folts

    Log entry time 20:19:45 on October 15, 2005

    Entry number 156968

    Followups:

    keyword=tech /right Dipole

    Dipole controls were stuck, I rebooted the Allen Bradley and quenched the magnet. in hind sight I should have tried a lead flow reset first.

    The first fault was a FC and a ZFCT, after the cryo recoverd I reset from the computer but the faults did not clear on the power supply but did on the computer. I tried to start the PS but got into chasing the phantom fault that leaves the on light on on the computer but no fault on the PS.

    Went through many iterations of this because in the past it would clear in time.

    The fault then changed to an FC and the computer no longer changed to on.

    The staus now is the FC on the power supply will not clear at all ( not even rebooting the AB) and the computer clears on the screen but faults when you try to start the PS.

    Jack says it may be in the firing circuit or the SCRs. We will wait till Mon. to start this work.