User name paschke
Log entry time 15:14:48 on October20,2005
Entry number 157505
Followups:
keyword=position difference stability
Chao's last PZT monitor scan has revealed that the matching solution he has
installed in not very stable... in particular, he expects a growth in the
horizontal position differences since the beginning of the run. He has no
information tracking changes on a daily basis, although such frequent
monitoring will start now.
The position differences observed in the Hall have been somewhat larger than
observed in the first slug, and this appears to be somewhat independent of
half-wave plate state. In Figure 1, I show position differences for the 5 slugs
(NOT sign corrected by waveplate state). Each slug represents data averaged
over an alternating waveplate state.
In principle the source terms could be changing for these plots, either
changing with halfwave plate state or drifting over time. We don't have good
information on that, since Hall B beam is on in the 100 keV region. While the
position differences for single slugs are larger than those measured in the
first slug, the statistical evidence for (or against) a clear change is weak.
As one would expect for statistical distributions around zero (or,
equivalently, for small but drifting distributions around zero), the
sign-corrected average is very small (after 5 slugs, nothing larger than 7 nm
+/- 5 nm).