    User name J. Singh

    Log entry time 01:44:50 on November 8,2005

    Entry number 159453

    keyword=problem with compton coincidence rates?

    during compton run 12752, for the last 45 minutes or so, the rate associated with the coincidences between the electron and photon detectors has "looked normal" but is not correlated at all with the cavity power (compton laser turning on and off). see fig 1 for the stripchart.

    red - cavity power
    blue - electron rate
    green - photon rate
    burgundy - coincidence rate

    see fig 2 for a zoomed in stripchart of when the problem started.

    normally, all the rates drop to near 0 when the laser is off to measure background.

    the run just ended. we'll start a new run after a coda reboot and see what happens