• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT

    User name lkaufman

    Log entry time 23:59:42 on November 9,2005

    Entry number 159648

    Followups:

    keyword=Cavity calibrations for slug 1: double-peaking of residuals

    To calibrate the position cavity monitors, I plot the predicted position 
    difference using the stripline data along the x axis (units are microns); and 
    along the y axis, I plot the measured position difference in the cavity (units 
    are channels).  The slope gives me the calibration constant in units of 
    chan/um, so I take the inverse of the slope as the calibration constant so that 
    I can convert the position differences measured by the cavities in channels to 
    microns.
    
    Both figures are only for slug 1: Figure 1 shows calibrations for cav2 and 
    figure 2 shows calibrations for cav3.  The top two plots in each figure show 
    the linear fit to the data from which I calculate the calibration constant for 
    the cavities.  The bottom two plots show the residual of the the fit in one 
    dimension.  The x axis of the residual plots is in units of channels.  It is 
    obvious from these plots that the fit for the y cavities (right column) is much 
    worse than for the x cavities (left column).  The residuals are not only wider 
    in y, but they are also double-peaked.  This double-peaking can also be seen in 
    the raw or regressed position differences as measured by the y cavities (shown 
    in halog 158096) and seems to be a problem with the cavities (and not the 
    striplines) based on the results I showed in halog 159607.
     
    I know that there is a lot going on here, and we are just beginning to try to 
    understand what's going on, so stay tuned.
    


    FIGURE 1

    FIGURE 2