Main INDEX
Monthly INDEX
PREV
NEXT
User name lkaufman
Log entry time 23:59:42 on November 9,2005
Entry number 159648
Followups:
keyword=Cavity calibrations for slug 1: double-peaking of residuals
To calibrate the position cavity monitors, I plot the predicted position
difference using the stripline data along the x axis (units are microns); and
along the y axis, I plot the measured position difference in the cavity (units
are channels). The slope gives me the calibration constant in units of
chan/um, so I take the inverse of the slope as the calibration constant so that
I can convert the position differences measured by the cavities in channels to
microns.
Both figures are only for slug 1: Figure 1 shows calibrations for cav2 and
figure 2 shows calibrations for cav3. The top two plots in each figure show
the linear fit to the data from which I calculate the calibration constant for
the cavities. The bottom two plots show the residual of the the fit in one
dimension. The x axis of the residual plots is in units of channels. It is
obvious from these plots that the fit for the y cavities (right column) is much
worse than for the x cavities (left column). The residuals are not only wider
in y, but they are also double-peaked. This double-peaking can also be seen in
the raw or regressed position differences as measured by the y cavities (shown
in halog 158096) and seems to be a problem with the cavities (and not the
striplines) based on the results I showed in halog 159607.
I know that there is a lot going on here, and we are just beginning to try to
understand what's going on, so stay tuned.
FIGURE 1
FIGURE 2