    Log entry time 02:45:35 on November 17, 2005

    Entry number 160405

    This entry is a followup to: 158096

    keyword=Cavity Monitor Resolution for Slug 13

    Here's a figure showing the resolution for the cavities in microns for
    slug 13. As a reminder, slug 13 is only different from slug 1 in that
    the cavity ioc was rebooted beforehand, and new phase offsets were set.
    This reboot got rid of the double-peaking previously seen in the position
    differences from the cavities. The resolution for all monitors is now
    1.3-1.8 um with the calibrations I calculated for this slug.




    Figure 1